Remote Control Commands
R&S
®
SMBV100B
872User Manual 1178.4460.02 ─ 03
Manual operation: See " Current Level" on page 328
[:SOURce<hw>]:POWer:MODE <Mode>
Selects the operating mode of the instrument to set the output level.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW and FIXed are synonyms.
To set the output level value, use the command [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude].
SWEep
Sets sweep mode.
Set the range and current level with the commands:
[:SOURce<hw>]:POWer:STARt and [:SOURce<hw>]:
POWer:STOP,
[:SOURce<hw>]:POWer:MANual.
*RST:
CW
Example: See Example "Setup an RF frequency or power sweep"
on page 882
Manual operation: See "State (RF level sweep)" on page 327
[:SOURce<hw>]:POWer:PEP?
Queries the PEP "Peak Envelope Power) of digital modulation or digital standards at
the RF output. This value corresponds to the level specification, displayed in the status
bar (header).
Return values:
<Pep> float
Example:
POW:PEP?
Response: "4"
The PEP value of digital modulation is 4 dBm at the RF output.
Usage: Query only
[:SOURce<hw>]:POWer:POWer <Power>
Sets the level at the RF output connector.
This value does not consider a specified offset.
The command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
sets the level of the "Level" display, that means the level containing offset.
See "RF frequency and level display with a downstream instrument" on page 288.