Remote Control Commands
R&S
®
SMBV100B
724User Manual 1178.4460.02 ─ 03
Example: See Example "Enable power ramping" on page 714
Manual operation: See "Ramp Time" on page 127
[:SOURce<hw>]:BB:DM:PRAMp:FDELay <FDelay>
[:SOURce<hw>]:BB:DM:PRAMp:RDELay <RDelay>
Sets the delay in the rising edge.
Parameters:
<RDelay> float
Range: 0 to 4
Increment: 0.01
*RST: 0
Default unit: symbol
Example: See Example "Enable power ramping" on page 714
Manual operation: See "Rise Delay" on page 128
[:SOURce<hw>]:BB:DM:PRAMp:ATTenuation <Attenuation>
Sets the level attenuation for signal ranges that are flagged with level attribute attenu-
ated by the control signal.
Parameters:
<Attenuation> float
Range: 0 to 50
Increment: 0.1
*RST: 15
Default unit: dB
Example: See Example "Enable power ramping" on page 714
Manual operation: See "Attenuation" on page 128
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe] <State>
Enables power ramping in the baseband only or mixed power ramping in the baseband
and the RF section.
The ON setting is mandatory if, with power ramping active, only the baseband signal is
output (I/Q outputs).
Only then can a signal with a defined, predictable level be output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Enable power ramping" on page 714.
Manual operation: See "In Baseband Only" on page 128