Remote Control Commands
R&S
®
SMBV100B
726User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth <SLength>
Defines the length of the signal sequence to be output in the SINGle trigger mode.
Parameters:
<SLength> integer
Range: 1 to 4294967295
*RST: 1000
Default unit: symbol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 713
Manual operation: See "Signal Duration" on page 111
[:SOURce<hw>]:BB:DM:TRIGger:RMODe?
Queries the status of signal generation.
Return values:
<RMode> STOP | RUN
Example:
SOURce1:BB:DM:TRIGger:SOURce ELTRigger
SOURce1:BB:DM:TRIGger:SEQuence ARETrigger
SOURce1:BB:DM:TRIGger:RMODe?
Response: RUN
Usage: Query only
Manual operation: See "Running/Stopped" on page 111
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
Enables signal output synchronous to the trigger event.
Parameters:
<Output> 0 | 1 | OFF | ON
*RST: 1
Example: See Example "Adjusting clock, marker and trigger settings"
on page 713
Manual operation: See "Sync. Output to External Trigger/Sync. Output to Trigger"
on page 112
[:SOURce<hw>]:BB:DM:TRIGger:ARM:EXECute
Stops signal generation; a subsequent internal or external trigger event restart signal
generation.
Example:
See also Example "Adjusting clock, marker and trigger settings"
on page 713
Usage: Event