EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 477

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
File and Data Management
R&S
®
SMBV100B
475User Manual 1178.4460.02 ─ 03
Figure 9-1: Parameter and operating modes, reset by the respective preset functions
Overview of the characteristics of the preset functions
Select the preset option that most fits to your particular application:
[Preset]
It is the most frequently used function.
A Preset executes a defined instrument setup to provide an initial instrument state
as a basis for a new configuration. It resets all parameters and switching states,
including also the states of inactive operating modes.
Network, remote access or system settings are retained.
To execute a preset, press the [Preset] key at the front panel.
"Set to Default"
Set To Default relates to individual dialogs or tabs and resets the associated set-
tings of the corresponding dialog. All other settings are retained.
For example, you can use it to reset all digital modulation settings in the "Custom
Digital Mod" dialog.
To reset the grouped settings, click "Set To Default".
"Preset this parameter"
Sets a single parameter to its default value.
Restoring the (Default) Instrument Configuration

Table of Contents

Related product manuals