Remote Control Commands
R&S
®
SMBV100B
715User Manual 1178.4460.02 ─ 03
Manual operation: See "State" on page 118
[:SOURce<hw>]:BB:DM:PRESet
Sets the default settings for digital modulation (*RST values specified for the com-
mands).
Not affected is the state set with the command SOURce<hw>:BB:DM:STATe
Example: See Example "Performing general tasks" on page 712
Usage: Event
Manual operation: See "Set To Default" on page 118
[:SOURce<hw>]:BB:DM:SRATe <SRate>
Sets the symbol rated in Hz/kHz/MHz or Symb/s / kSymb/s and MSymb/s.
Parameters:
<SRate> float
Range: 400 to depends on the installed options
Increment: 0.001
*RST: 270833.333
Default unit: Hz
Example: See Example "Performing general tasks" on page 712.
Manual operation: See "Symbol Rate" on page 119
[:SOURce<hw>]:BB:DM:STANdard <Standard>
Selects predefined set of settings according to the selected standard, see Table 3-6.
Parameters:
<Standard> USER | BLUetooth | DECT | ETC | GSM | GSMEdge | NADC |
PDC | PHS | TETRa | W3GPp | TDSCdma | CFORward |
CREVerse | WORLdspace | TFTS | APCOPH1C4fm |
APCOPH1CQpsk | APCOPH2HCpm | APCOPH2HDQpsk |
APCOPH2HD8PSKW | APCOPH2HD8PSKN | APCOPH1Lsm |
APCOPH1Wcqpsk
A query returns the value USER if one the following is true:
A user-defined custom digital modulation setting was loaded
One of the associated settings was changed subsequent to the
selection of a standard.
*RST: GSM
Example: See Example "Performing general tasks" on page 712
Manual operation: See "Set according to Standard" on page 118
SOURce Subsystem