Remote Control Commands
R&S
®
SMBV100B
719User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:DM:FILTer:PARameter:RCOSine <RCosine>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:SPHase <SPhase>
Sets the filter parameter.
Filter Type Parameter Parameter
Name
Min Max Increment Default
APCO25
Roll-off factor
<Apco25>
0.05 0.99 0.01 0.2
APCO25Lsm
Cut off frequency for the
lowpass/gauss filter
(:LOWPass/:GAUSs)
<Cosine>
400 25E6 1E-3 270833.333
COSine
Bandwidth
<FiltParm>
400 depends on the installed
options
*)
1E-3 270833.333
COSine
Roll-off factor
<Cosine>
0.05 1 0.01 0.35
GAUSs
Roll-off factor
<Gauss>
0.15 100000 0.01 0.3
LPASs
Cut-off frequency
<LPass>
0.05 2 0.01 0.5
LPASSEVM
Cut-off frequency
<LPassEvm>
0.05 2 0.01 0.5
PGAuss
Roll-off factor
<PGauss>
0.15 2.5 0.01 0.3
RCOSine
Roll-off factor
<RCosine>
0.05 1 0.001 0.35
SPHase
B x T
<SPhase>
0.15 2.5 0.01 2
*)
100E6 (base unit) / 200E6 (R&S SMBVB-K523) / 300E6 (R&S SMBVB-K524)
Parameters:
<SPhase> float
Range:  0.15  to  2.5
Increment:  0.01
*RST:  2
Example: 
SOURce:BB:DM:FILTe:TYPE SPHase
SOURce:BB:DM:FILTer:PARameter:SPHase 0.5
Manual operation:  See "Filter Parameter" on page 125
[:SOURce<hw>]:DM:FILTer:PARameter <Parameter>
Sets the filter parameter of the currently selected filter type.
To set the filter type, use command [:SOURce<hw>]:BB:DM:FILTer:TYPE
on page 718.
Parameters:
<Parameter> float
Range:  0.05  to  2.5
Increment:  0.01
*RST:  0.35
SOURce Subsystem