EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 744

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100B
742User Manual 1178.4460.02 ─ 03
SOURce:BB:ARBitrary:STATe ON
// ******************************************************************
// Select the next segment trigger mode and source
// ******************************************************************
SOURce:BB:ARBitrary:TRIGger:SMODe NEXT
// SOURce:BB:ARBitrary:TRIGger:SMODe NSEam
SOURce:BB:ARBitrary:WSEGment:NEXT:SOURce INTernal
SOURce:BB:ARBitrary:WSEGment?
// 0
SOURce:BB:ARBitrary:WSEGment:NAME?
// "/var/user/ARB/multi_segment/Seg_0.wv"
// Seg_0 is output continuously
// ******************************************************************
// Trigger a switch over to the next segment
// ******************************************************************
SOURce:BB:ARBitrary:WSEGment:NEXT 3
// stops Seg_0 and starts immediatly Seg_3; Seg_3 is output continuously
SOURce:BB:ARBitrary:TRIGger:EXECute
//restarts Seg_3
SOURce:BB:ARBitrary:WSEGment:NEXT 2
SOURce:BB:ARBitrary:WSEGment:NEXT 0
// ******************************************************************
// Scrolling trough the segments, i.e. shwitch over in incremental order
// ******************************************************************
SOURce:BB:ARBitrary:WSEGment:NEXT:EXECute
SOURce:BB:ARBitrary:WSEGment?
// 1
SOURce:BB:ARBitrary:WSEGment:NEXT:EXECute
SOURce:BB:ARBitrary:WSEGment?
// 2
Refer to "To apply a manual segment switch to the output segments in the required
order" on page 216 for description of the steps necessary to achieve a similar task via
manual operation of the instrument.
Example: Using the ARB sequencer
The following example lists the commands necessary to configure a play list. The
example lists only the relevant commands.
We assume, that the multi-segment sequence ms_0to3 composed of four segments,
Seg_0, Seg_1, Seg_2 and Seg_3 is created and loaded in the ARB (see Exam-
ple "Creating a multi-segment waveform" on page 740). The required output order of
the segments is Seg_0, Seg_3 [2], Seg_2.