EasyManua.ls Logo

Inficon HAPSITE Smart - Page 307

Inficon HAPSITE Smart
554 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 - 21
IPN 074-397-P1G
HAPSITE Smart Operating Manual
3 Enter the name of the compound, to be located, in the box on the top left of the
screen. EXAMPLE: Benzene. See Figure 9-23.
Figure 9-23 NIST Name Entry
4 The spectrum and information about the compound will appear in the two boxes
to the right of the names column.
5 In the bottom right box, the 10 Largest Peaks will be listed. Make a note of the
three largest mass peaks that are between 45-300 AMU.
HINT: Peaks are listed in order from the largest to the smallest. EXAMPLE:
Benzene's 3 largest peaks are masses 78, 77 and 51. See Figure 9-24.
Figure 9-24 Top 10 Masses
6 Minimize the NIST window and return to the Smart IQ Data screen displaying
the TIC chromatogram.
7 Enter the 3 largest peaks, from the compound in step 5, in the control panel
box. Enter each mass number in the boxes under the Ma... (Mass) column. See
Figure 9-20.
8 Check the box to the left of the mass or press ENTER. This will plot the mass
in the RIC window and create a new row in the control panel for an additional
mass to be entered. See Figure 9-25.
Figure 9-25 Entering Masses in Control Panel
9 As soon as the first mass is checked, a RIC plot will open directly under the TIC
plot. See Figure 9-26.
NOTE: This RIC window can be closed by un-checking the masses selected in the
control panel.

Table of Contents

Related product manuals