15 - 10
IPN 074-397-P1G
HAPSITE Smart Operating Manual
Figure 15-2 Moving Methanol Around in Probe and Probe Line
5 Empty the methanol remaining in the probe line by tipping one end of the probe
line downward until all of the methanol drains from the probe line. See Figure
15-3.
Figure 15-3 Empty Methanol from Probe and Probe Line
6 Blow out the probe line with Nitrogen to remove any residual methanol that may
be left in the probe line. See Figure 15-4.