EasyManua.ls Logo

Inficon HAPSITE Smart - Page 468

Inficon HAPSITE Smart
554 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15 - 10
IPN 074-397-P1G
HAPSITE Smart Operating Manual
Figure 15-2 Moving Methanol Around in Probe and Probe Line
5 Empty the methanol remaining in the probe line by tipping one end of the probe
line downward until all of the methanol drains from the probe line. See Figure
15-3.
Figure 15-3 Empty Methanol from Probe and Probe Line
6 Blow out the probe line with Nitrogen to remove any residual methanol that may
be left in the probe line. See Figure 15-4.

Table of Contents

Related product manuals