No. 99MBB765A
XYZ Adjustment Table screen
Touch .
XYZ Adjustment Table screen
Turn the Y-
axis adjustment knob to search for the maximum
position of the Z coordinate, and touch .
XYZ Adjustment Table screen
Enter the micrometer head value of the Y-axis adjustment knob,
and touch
.
XYZ Adjustment Table screen
Touch [Drive Unit Operation] to display the Drive Unit
Operation screen.
Drive Unit Operation screen
Move the stylus to the second point.
Touch the buttons according to the direction and distance to
reposition the stylus.
The movement speed differs depending on the size of the button
you press. The detector moves fastest with the largest button. As
you touch the smaller buttons, the detector moves slower.
Touching moves the stylus toward the origin position.
Touching
moves the stylus away from the origin position.
11 Touch [OK].