CON command
Command to modify measurement/evaluating conditions
Data section Bytes: number of bytes from the tip of data
0: A conditions, 1: B conditions
0: JIS1982, 1: JIS1994, 2: JIS2001, 3: ISO1997, 4: ANSI, 5: VDA, 6: FREE
0: P, 1: R, 2: DF, 3: R.MOTIF, 4: W.MOTIF, 5: W
0: 2CR75, 1: PC75, 2: GAUSS, 3: None
0: 2.5 um (98.425 uin), 1: 8 um (314.960 uin), 2: 25 um (984.251 uin) 3: None
* (Cutoff length λc,
Sampling length L)
0: 0.08 mm (0.003 in), 1: 0.25 mm (0.009 in), 2: 0.8 mm (0.031 in),
3: 2.5 mm (0.098 in), 4: 8 mm (0.314 in), 5: 25 mm ( 984.251 uin)
* The command is ignored when R/W.MOTIF profile is selected.
0: 0.25 mm (0.009 in), 1: 0.8 mm (0.031 in), 2: 2.5 mm (0.098 in),
3: 8 mm (0.314 in), 4: NONE
* The command is ignored when the item other than W profile is selected.
** (Number of sampling lengths)
00 - 20 (00 is designate arbitrary length.)
* The command is ignored when R/W.MOTIF profile is selected.
0.1 - 50.80
(effective when the number of sampling length is "00")
* (Upper limit of motif length A)
0: Arbitrary length, 1: 0.02 mm (0.0 in), 2: 0.1 mm (0.003 in),
3: 0.5 mm (0.019 in), 4: 2.5 mm (0.098 in)
* The command is ignored when the item other than R/W.MOTIF profile is selected.
* (Upper limit of motif length B)
0: Arbitrary length, 1: 0.1mm (0.003 in), 2: 0.5mm (0.019 in),
3: 2.5 mm (0.098 in), 4: 12.5 mm (0.492 in)
* The command is ignored when the item other than W.MOTIF profile is selected.
**.**
(Arbitrary maximum length A)
0.01 - 50.79 mm (0.001 - 1.999 in)
* The command is ignored when the item other than R/W.MOTIF profile is selected.
**.**
(Arbitrary maximum length B)
0.01 - 50.79 mm (0.001 - 1.999 in)
* The command is ignored when the item other than W.MOTIF profile is selected.
***: Refer to “● Error codes”.
**:Bytes with error code