Triggers
R&S
®
RTP
262User Manual 1337.9952.02 ─ 12
Length ← Frame Id
Sets the length of the pattern for the trigger alignment condition.
Remote command:
TRIGger<m>:SERPattern:COMPlex:SOFLength on page 1285
Bit Pattern ← Frame Id
Sets the bit pattern value for the trigger alignment condition.
Remote command:
TRIGger<m>:SERPattern:COMPlex:SOFPattern on page 1285
Timeout ← Frame Id
Sets the timeout, a duration that defines an idle mark.
Remote command:
TRIGger<m>:SERPattern:COMPlex:SOFTimeout on page 1284
Idle range
Selects if the timeout should be within the range or out of range.
Remote command:
TRIGger<m>:SERPattern:COMPlex:IDLRange on page 1284
Logic
Selects the logic for the condition, if more than one pattern conditions are defined.
"OR"
At least one of the defined pattern conditions have to be met.
"AND"
All the defined pattern conditions have to be met.
Remote command:
TRIGger<m>:SERPattern:COMPlex:CLOGic on page 1283
Bit order
Select most significant bit (MSB) or least significant bit (LSB) to define the data bit
order.
Remote command:
TRIGger<m>:SERPattern:COMPlex:BORDer on page 1283
Bit offset
Sets the number of arbitrary bits between the first bit after the frame id and the first bit
of the first pattern condition to be triggered on.
Remote command:
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:BOFFset on page 1285
Pattern length
Sets the word pattern length in bits. The maximum bit length of all selected pattern
conditions is 160 bits, with arbitrary bits (e.g., due to bit offsets) not counted.
Remote command:
TRIGger<m>:SERPattern:COMPlex:PATTern<n>:PATLength on page 1286
High-speed serial pattern trigger (Option R&S