Configuring the Internal Baseband Source
R&S
®
SMBV100B
240User Manual 1178.4460.02 ─ 03
6. Select "ARB: Multi Carrier > Carrier Graph" to visualize the configuration.
7. Select "ARB: Multi Carrier > General > Output Settings > Output File" and enter the
name of the output file, for example 4x5MHz_3gpp_1_16.
8. To trigger signal calculation and to load the waveform in the ARB generator, select
"ARB: Multi Carrier > General > Create/Create and Load" function.
9. Select "ARB > General > Load Waveform".
Confirm that the generated multi-carrier waveform is used.
10. Select "ARB > State > On" to enable the ARB generator.
Generating Multi-Carrier Signals