File and Data Management
R&S
®
SMBV100B
479User Manual 1178.4460.02 ─ 03
Table 9-1: Key parameters affected by preset and factory preset
Parameter Preset value Preset Factory
Preset
RF frequency 1 GHz x x
RF level (RF output) off x x
RF OFF mode - - x
Offsets 0 x x
Modulation state off x x
Uninterrupted level settings off x x
Level attenuator mode auto x x
Level ALC (internal level control) auto x x
Level UCOR (user correction) off x x
LF output state off x x
Sweep state off x x
List mode state off x x
Reference frequency settings (reference oscillator) - - x
Power on settings (Level/EMF) - - x
Network settings - - x
Hostname - - x
GPIB address - - x
Start/Stop display update - - x
Display and keyboard settings - - x
Preset behavior: Keep connections to external instru-
ments
off - x
Password and settings protected by passwords (e.g. dis-
abled LAN or USB)
- - -
Security settings - - -
User files (setups
2)
, data lists, etc.)
- - -
●
2)
 UserPreset.savrcltxt is renamed as UserPresetInactive.savrcltxt;
an existing file with the same name is overwritten.
If the default values in the "Remote Access" dialog had been changed, a factory preset
via remote control (:SYSTem:FPReset) terminates the connection to the instrument.
Security settings are never reset.
Resets all parameters and switching states, and closes all opened dialogs.
Restoring the (Default) Instrument Configuration