Remote Control Commands
R&S
®
SMBV100B
616User Manual 1178.4460.02 ─ 03
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Example:
DEV:PRES
Presets all instruments settings that are not related to the signal
path.
Usage: Event
Manual operation: See "Connect/Disconnect All Remote" on page 446
:SOURce<hw>:PRESet
Presets all parameters which are related to the selected signal path.
Example:
SOUR:PRES
Presets all settings that are related to signal path
Usage: Event
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
●
The [Preset] key
●
The *RST command
For an overview of the settings affected by the preset function, see Table 9-1
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage: Setting only
:SYSTem:FPReset
Triggers an instrument reset to the original state of delivery.
Example:
SYST:FPR
All instrument settings (also the settings that are not currently
active) are reset to the factory values.
Usage: Event
Manual operation: See "Execute Factory Preset" on page 477