EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 448

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi-Instrument Setups
R&S
®
SMBV100B
446User Manual 1178.4460.02 ─ 03
Connect/Disconnect All Remote
Triggers the instrument to:
Establish the remote connections to all configured External Instruments.
To terminate all existing remote connections.
The external instruments are not removed from the list of external instruments.
Only the remote connection status changes.
Remote connection status can be toggled with the parameter Remote Connection Sta-
tus.
Note:
External instruments connected to the R&S SMBV100B are preset (:DEVice:PRESet
) whenever:
The connection to the external instrument is established or terminated ("Connect/
Disconnect All Remote" or "Rem. Conn > On/Off")
The R&S SMBV100B is preset ([Preset]).
To ensure that the remote connections to the external instruments will be automatically
established after instrument's start-up or a reboot, set Auto Connect > "On".
Remote command:
:SCONfiguration:EXTernal:REMote:CONNect[:ALL] on page 667
:SCONfiguration:EXTernal:REMote:DISConnect[:ALL] on page 667
Preset behavior: Keep connections to external instruments
Sets the effect of R&S SMBV100B preset ([Preset] or *RST) on the following:
The connection status between the instrument and the connected external instru-
ments.
The R&S SMBV100B itself and the preset values.
The connected instruments.
"Preset behavior..." Off On
R&S SMBV100B and preset
values
Standard R&S SMBV100B preset
See Table 9-1.
Standard R&S SMBV100B preset
Excluded from preset:
"I/Q Output Type"
See also
Connection status Terminated ("Disconnect All Remote") Unchanged
Connected instruments External instrument preset (:DEVice:PRESet) External instrument preset
Remote command:
:SCONfiguration:EXTernal:PBEHaviour on page 667
Connector Name
Displays the connector name.
Remote command:
via keyword in the SCPI syntax, e.g. IQOutput
Direction
Displays the connector direction as a color coded field.
System Configuration Settings

Table of Contents