Multi-Instrument Setups
R&S
®
SMBV100B
447User Manual 1178.4460.02 ─ 03
Remote command:
:SCONfiguration:EXTernal:DIGital:DIRection? on page 668
:SCONfiguration:EXTernal:IQOutput:DIRection? on page 668
External Instrument
Accesses a dialog with settings that help you to establish the connection to the exter-
nal instruments, see Chapter 8.5.2.2, "External Instrument Configuration Settings",
on page 449.
Note:
Each external instrument must be connected via a LAN or USB to the
R&S SMBV100B.
In most test setups, all instruments must use a common reference frequency signal,
either from an external common reference source or using the distributed reference
frequency of the R&S SMBV100B.
I/Q Connection
Depends on the interface type.
●
Analog I/Q interfaces:
Selects the type of the I/Q output signal
●
Digital I/Q interfaces:
Indicates the status of the I/Q connection of the digital interfaces.
The indication includes a status LED and an icon, where the icon depends on the
HW connector (Dig I/Q or HS Dig I/Q) used.
Indication Designation
A green LED and an
icon
The Dig I/Q interface is used as an input connector
A yellow LED and an
icon
The Dig I/Q/HS Dig I/Q interface is used as an output connector
An icon without an LED An external instrument is connected to the Dig I/Q interface, but no direction is
specified
No icon displayed No connection
"Single Ended"/"Differ-
ential"
Selects the type of the I/Q output signal at the I/Q output connector
Works like the parameter "I/Q Output Type" on page 468.
Remote command:
:SCONfiguration:EXTernal:DIGital:IQConnection:STATe? on page 668
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE on page 835
Remote Connection Status
Indicates the status of the remote connection. Click the status indicator to toggle the
remote connection status.
To connect/disconnect all external instruments, use the functions Connect/Disconnect
All Remote.
Note: