Remote Control Commands
R&S
®
SMBV100B
835User Manual 1178.4460.02 ─ 03
Setting parameters: 
<Filename> "<filename>"
Filename or complete file path
Usage:  Setting only
Manual operation:  See "Save/Recall" on page 467
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE <Type>
Sets the type of the analog signal.
Parameters:
<Type> SINGle | DIFFerential
*RST:  SING (if SCONfiguration:EXTernal:PBEHaviour 0) /
- (if SCONfiguration:EXTernal:PBEHaviour 1)
Example: 
SOURce1:IQ:OUTPup:ANALog:TYPE DIFFerential
Options:  DIFFerential requires R&S SMBVB-K17
Manual operation:  See "I/Q Connection" on page 447
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:MODE <Mode>
Determines the mode for setting the output parameters.
Parameters:
<Mode> FIXed | VARiable | VATTenuated
FIXed
Locks the I/Q output settings
VARiable
Unlocks the settings
VATTenuated
Attenuates the signal with the value set with the command [:
SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:
DATTenuation.
*RST:  FIXed
Example:  See Example "Enabling and configuring the analog output"
on page 833.
Manual operation:  See "Mode" on page 468
[:SOURce<hw>]:IQ:OUTPut:LEVel <Level>
Sets the off-load voltage Vp of the analog I/Q signal output.
The value range is adjusted so that the maximum overall output voltage does not
exceed 4V, see "Maximum overall output voltage" on page 466.
SOURce Subsystem