Remote Control Commands
R&S
®
SMBV100B
656User Manual 1178.4460.02 ─ 03
Example:
SENSe1:POWer:CORRection:SPDevice:STATe 1
Activates the use of the S-parameters correction data.
Manual operation: See " S-Parameter " on page 402
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been loaded to the power sen-
sor.
Return values:
<List> string list
*RST: 0
Usage: Query only
Manual operation: See " S-Parameter " on page 402
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority <Priority>
Selects average or peak power for permanent display.
Parameters:
<Priority> AVERage | PEAK
*RST: AVERage
Example:
SENS1:DISP:PERM:STAT ON
The permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
The measured average power is indicated.
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe <State>
Activates the permanent display of the measured power level results. The instrument
also indicates the sensor type, the connection, the measurement source and the offset
if set.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS1:POW:DISP:PERM:STAT ON
The permanent viewer is switched on.
Manual operation: See " Permanent " on page 400
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
Queries the current filter length in filter mode AUTO (:SENSe<ch>[:POWer]:
FILTer:TYPE)
SENSe, READ, INITiate and SLISt Subsystems