Remote Control Commands
R&S
®
SMBV100B
655User Manual 1178.4460.02 ─ 03
:SENSe<ch>[:POWer]:APERture:DEFault:STATe <UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command :SENSe<ch>[:POWer]:
APERture:TIMe on page 655.
Parameters:
<UseDefAp> 0 | 1 | OFF | ON
*RST: 1
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 651.
Manual operation: See "Default Aperture Time" on page 402
:SENSe<ch>[:POWer]:APERture:TIMe <ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
Parameters:
<ApTime> float
Range: depends on connected power sensor
Increment: 1E-9
*RST: depends on connected power sensor
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 651.
Manual operation: See "Aperture Time" on page 402
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect <Select>
Several S-parameter tables can be stored in a sensor. The command selects a loaded
data set for S-parameter correction for the corresponding sensor.
Parameters:
<Select> float
*RST: 0
Manual operation: See " S-Parameter " on page 402
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe <State>
Activates the use of the S-parameter correction data.
Note: If you use power sensors with attenuator, the instrument automatically activates
the use of S-parameter data.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0