Remote Control Commands
R&S
®
SMBV100B
747User Manual 1178.4460.02 ─ 03
Parameters:
<TSig> float
Range: -1 to 1
Increment: 0.001
*RST: 0
Default unit: FS
Example: See Example "Creating test signals" on page 737
Manual operation: See "I Value, Q Value" on page 158
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:AMPLitude
<Amplitude>
Sets the digital amplitude of the rectangular wave.
Parameters:
<Amplitude> float
Range: 0 to 1
Increment: 0.001
*RST: 0.800
Default unit: FS
Example: see Example "Creating test signals" on page 737
Manual operation: See "Amplitude" on page 157
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:FREQuency <Frequency>
Sets the frequency of the test signal.
Parameters:
<Frequency> float
Range: 100 to depends on the installed options
Increment: 0.01
*RST: 1000
Default unit: Hz
Example: See Example "Creating test signals" on page 737.
Manual operation: See "Frequency" on page 157
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:OFFSet <Offset>
Sets the DC component.
Parameters:
<Offset> float
Range: -1 to 1
Increment: 0.001
*RST: 0
Default unit: FS
SOURce Subsystem