Remote Control Commands
R&S
®
SMBV100B
786User Manual 1178.4460.02 ─ 03
Parameters:
<CFactor> float
Range: 0 to 30
Increment: 0.01
*RST: 3
Example: See "Programming Examples" on page 779
Manual operation: See "Desired Crest Factor" on page 244
[:SOURce<hw>]:BB:MCCW:CFACtor:ACTual?
Queries the actual Crest Factor for optimization mode target crest.
Return values:
<Actual> float
Range: 0 to 100
Increment: 0.01
*RST: 3
Example: See "Programming Examples" on page 779
Usage: Query only
[:SOURce<hw>]:BB:MCCW:CARRier:STATe <CarrierIndex>, <State>
Switches the selected carrier on or off.
Parameters:
<CarrierIndex> integer
Range: 0 to lastCarrier
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See "Programming Examples" on page 779
Manual operation: See "Carrier Table" on page 246
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe <Stat0[,Stat1..]>]
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe? <Start>, <Count>
Switches the carrier on or off with the aid of a value list.
The first value in the list is assigned to the carrier with index 0, the second value to the
carrier with index 1, etc. The maximum length corresponds to the maximum number of
multi carriers. There is no need to enter all the values every time. Values not set by the
value list are set with the default values provided they have already been explicitly set
by a previous command. If this is the case, the values continue to apply until overwrit-
ten.
Setting parameters:
<Stat0[,Stat1..]> ON | OFF
SOURce Subsystem