Remote Control Commands
R&S
®
SMBV100B
815User Manual 1178.4460.02 ─ 03
Example:  See Example "Create a table with user-defined correction values
for the RF level" on page 811 .
Usage:  Query only
Manual operation:  See "User Correction" on page 385
[:SOURce<hw>]:CORRection:ZERoing:STATe <State>
Activates the zeroing procedure before filling the user correction data acquired by a
sensor.
Parameters:
<State> 0 | 1 | OFF | ON
*RST:  1
Example:  See Example "Fill user correction data with sensor"
on page 812.
Manual operation:  See " Fill User Correction Data with Sensor" on page 389
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction list with the level values measured by the power sen-
sor for the given frequencies.
Suffix: 
SENSor<ch>
.
Defines the used power sensor, i.e. the sensor whose values are
used.
Example:  See Example "Fill user correction data with sensor"
on page 812.
Usage:  Event
[:SOURce<hw>]:CORRection[:STATe] <State>
Activates user correction with the currently selected table.
Parameters:
<State> 0 | 1 | OFF | ON
*RST:  0
Example:  See Example "Create a table with user-defined correction values
for the RF level" on page 811 .
Manual operation:  See "State" on page 385
[:SOURce]:CORRection:CSET:CATalog?
Queries a list of available user correction tables.
SOURce Subsystem