Remote Control Commands
R&S
®
SMBV100B
857User Manual 1178.4460.02 ─ 03
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: depends on the installed options
Default unit: dBm
Example: See Example "Create an RF list and activate the list mode"
on page 850.
Manual operation: See "Edit List Mode Data" on page 343
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 850.
Usage: Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:
See Example "Create an RF list and activate the list mode"
on page 850.
Usage: Event
Manual operation: See " Execute Single" on page 338
[:SOURce<hw>]:LIST:TRIGger:SOURce <Source>
Selects the trigger source for processing lists.
The names of the parameters correspond to those in sweep mode. SCPI standard
uses other names for the parameters; these names are also accepted by the instru-
ment. The SCPI names should be used if compatibility is an important consideration.
For an overview, see the following table:
SOURce Subsystem