EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 889

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100B
887User Manual 1178.4460.02 ─ 03
LINear
Sets a fixed frequency value as step width and adds it to the cur-
rent frequency.
The linear step width is entered in Hz, see [:SOURce<hw>]:
SWEep[:FREQuency]:STEP[:LINear].
LOGarithmic
Sets a constant fraction of the current frequency as step width
and adds it to the current frequency.
The logarithmic step width is entered in %, see [:
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic.
*RST: LINear
Example: See Example "Setup an RF frequency or power sweep"
on page 882.
Manual operation: See " Spacing" on page 330
[:SOURce<hw>]:SWEep:POWer:SHAPe <Shape>
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe <Shape>
Determines the waveform shape for a frequency sweep sequence.
Parameters:
<Shape> SAWTooth | TRIangle
*RST: SAWTooth
Example: See Example "Setup an RF frequency or power sweep"
on page 882.
Manual operation: See " Shape" on page 329
[:SOURce<hw>]:SWEep:POWer:EXECute
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Executes an RF frequency sweep.
The command performs a single sweep and is therefore only effective in manual
sweep mode.
Example:
See Example "Setup an RF frequency or power sweep"
on page 882.
Usage: Event
Manual operation: See " Execute Single Sweep " on page 331
[:SOURce<hw>]:SWEep:POWer:RETRace <State>
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace <State>
Activates that the signal changes to the start frequency value while it is waiting for the
next trigger event.
SOURce Subsystem

Table of Contents

Related product manuals