Configuring the RF Signal
R&S
®
SMBV100B
329User Manual 1178.4460.02 ─ 03
Retrace
For "Shape = Sawtooth" and "Mode = Single/External Single", enables changing the
signal to the start value while it is waiting for the next trigger event. It returns to the
start level or start frequency until the next trigger is applied.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace on page 887
[:SOURce<hw>]:SWEep:POWer:RETRace on page 887
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RETRace on page 848
Shape
Selects the waveform shape of the sweep signal.
"Sawtooth"
The sweep runs from start to stop frequency. The subsequent sweep
starts at the start value, i.e. the shape of the sweep sequence resem-
bles a sawtooth.
Figure 6-16: Sweep signal sawtooth shape
1 = Start value
2 = Stop value
3 = Step width