Chapter 5
5.4 Exposure with CR cassette
2
Read the bar code of the CR cassette
with a bar code reader.
The read bar code number is displayed in the
Exam. Tags display area, and focus moves to
the next Exam. Tag. This completes the prepa-
rations for reading by REGIUS.
3
Perform exposure using the registered CR
cassette.
Configure the X-ray device settings and check
the exposure setting before making an expo-
sure.
Press the exposure switch
4
Read CR cassette.
The image read from the registered CR cassette
will be displayed.
When exposing several body parts, repeat from
step 1.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
For bar code registration, the optional bar code reader
is required.
The bar code reader is a CDRH class II laser product.
Operate it with caution on the following points.
Do not look into the front reading window. There is
a danger that the laser will damage the eyes.
Do not discharge the laser in the direction of
people's eyes. There is a danger that the laser will
damage the eyes.
Sometimes the bar code cannot be read correctly
when the laser strikes it perpendicularly. If it cannot be
read correctly, tilt the bar code reader slightly and read
again.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
X-ray exposure information is acquired automatically
when linked to the X-ray device.
When reading from the bar code of the CR cassette
that does not match the exposure conditions, a conr-
mation screen is displayed which says, "This cassette
is for general exposure. Check the Exam. Tag." Press
[OK], and register the correct CR cassette.
CR cassettes with the same bar code cannot be reg-
istered consecutively. When there are several expo-
sures with CR cassettes with the same bar code, re-
register after reading the image with the REGIUS for
each exposure.
By continually carrying out steps 1 and 2, the CR cas-
settes can be continually registered.
Exposure with bar code registration is suitable when
this device is connected with multiple REGIUS series
devices.
Congure irradiation elds of at least 5x5 cm for the
following areas.
Infant hip/(C) Infant hip/Front view of lumbar/Side view
of lumbar/Radial position of lumbar/Front view of head/
Towne view of the head/Turkey saddle/Cheekbone
axis/Temporomandibular joint/Front view of jawbone/
Side view of jawbone/Front view of the child chest/
Side view of the child chest/Progress observation of
child chest/Front view of infant chest/Infant chest and
abdomen/(C) Side view of infant chest and abdomen/
Progress observation of infant chest and abdomen/
Newborn chest/(C) Newborn chest/Newborn skeleton/
(C) Newborn skeleton/Front view of cervix/Side view
of cervix/Radial position of cervix/Cervix opening/
Ribs/Side view of sternum/Radial position of sternum/
Clavicle/Shoulder joint/Scapula/Sternoclavicular joint/
Hips/Limbs (excluding radial position of heel bone)
Congure settings for other areas so that at least one
edge of the irradiation eld is at least 15 cm.
If these conditions are not satised, irradiation elds
sometimes cannot be detected, which would require
reimaging.
•••••••••••••••••••••••••••••••••••••••••••••••••••••