338
The automatic emergency number issuing function au-
tomatically issues a patient ID and accession number
for the emergency patient who has been received in
order for the examination to proceed immediately.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When the checkup mode is turned ON, the automatic
emergency number issuing function cannot be used.
Turn the checkup mode OFF.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Press [trauma] on the initial screen.
• The body part selection screen is displayed.
• Check that the automatic issuing worked cor-
rectly.
HINT
•••••••••••••••••••••••••••••••••••••
• If [trauma] button is not displayed on the initial screen,
set it using the user tool.
• When automatic transition is turned OFF, the emer-
gency number is automatically issued from the initial
screen. Press [Perform] to start examination.
• If you have set to register the Exam. Tags automati-
cally, the body part selection screen is not displayed.
The display transitions to the exposure screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
7.19 Automatic emergency number issuing
2
Select the body part and press [Perform].
• The exposure screen is displayed.
HINT
•••••••••••••••••••••••••••••••••••••
• Press [Back] on the Body part selection screen to re-
turn the display to the initial screen. In this case, the
issued number remains the same.
• If you press [Schedule] on the Body part selection
screen, the number is saved and the display returns
to the initial screen. In this case, a new number is not
issued.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Perform exposure.
• Configure the X-ray device settings and check
the exposure setting before making an expo-
sure.
Press the exposure switch