1.2 Safety precautions
WARNING
Be sure to implement the following cyber-security
measures:
Provide for an adequate level of information se-
curity measures within the facility using this de-
vice, and other facilities connected to the facility
via networks.
Isolate the device from the connected network
immediately upon infection by malware, concur-
rently prohibiting the use of the infected device.
Upon infection of the device by malware, or
the potential infection thereof, promptly contact
Konica Minolta service representative.
Manage the account information (user name, ID,
etc.), password and other sensitive information as
follows:
Use a combination of upper- and lower-case al-
phabet and numeric characters to prevent easy
identication by malicious parties.
Do not leave records of sensitive information to
avoid uncertied disclosure.
Passwords should consist of at least eight char-
acters.
Passwords should be assigned with expiration
dates, and changed on a regular basis.
1.2.8 Precautions for each
function
WARNING
Precautions on stitching combination processing
Adjust the combination position by using the mark-
ers as supplementary standards in addition to body
parts of the subject. It is also recommended that
you arrange the measure in the combined image
part during exposure to simplify the conrmation of
the combined parts.
Precautions for One Shot stitching exposure
One Shot stitching exposure outputs the combined
picture by correcting the images' overlapping part
captured by DR panel. Depending on the capturing
conditions, lateral stripes or decrease in contrast
might occur in a correction area, or decrease in
contrast might also occur if a structure horizontally
long overlaps with the correction area. Make sure
the correction area and the parts in an examination
are not overlapping.
Make sure to observe the number of DR panels
that are loaded in the stitching holder through the
exposure screen before conducting the exposure.
Precautions for image measurements
Sometimes the exposure orientation causes ac-
cidental errors with regard to the actual length of
the subject. Use while taking into account any error
between image and measured results.
Precautions when running HS processing
The noise is reduced by HS processing, and you can
expect signals that are unclear due to noise to become
clearer, but signals that do not exist will not be visible.
When reviewing whether to reduce dosage by means
of HS processing, consider lowering the signal com-
ponent by means of reducing dosage and always
check whether images are suitable for diagnosis
.

Table of Contents

Other manuals for Konica Minolta CS-7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-7 and is the answer not in the manual?

Konica Minolta CS-7 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-7
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals