274
7.9 SIGMA stitching exposure
7.9.1 Screen structure
z
Exposure screen
This is the screen for SIGMA stitching exposure using
Reina buckies.
(7)
(3)
(1)
(6)
(8)
(5)
(4)
(9)
(2)
(10)
Number Name Functions
(1) Order list List of exposures
(2)
Combined image
display area
The images for combination
are displayed. Focus is dis-
played on selected images.
(3)
Standard grada-
tion button
This button is for select-
ing the image which will be
the image shade standard
among the images displayed
in the combined image
display area. You can only
select 1 image.
: Not selected
: Selected
(4)
Selection button
(batch selection)
Gathers multiple exposure
images associated with the
images selected in the order
list and displays them in the
combined image display
area.
(5)
Selection button
(individual selec-
tion)
Displays the selected images
in the order list in the com-
bined image display area.
(6) Delete button
Deletes the images selected
in the combined image dis-
play area.
(7)
Combination
sequence switch
button
Moves the position of the
images selected in the com-
bined image display area.
(8) Stitching button
Combines the images dis-
played in the combined im-
age display area. Press this
button to display the stitching
combination position adjust-
ment screen.
Number Name Functions
(9) Adjust button
Performs re-adjustment
of the combination posi-
tion. This button cannot be
pressed when combined
images are not displayed.
(10) SID setting Select the SID value.
HINT
•••••••••••••••••••••••••••••••••••••
• Adjustments to the combination position of the stitch-
ing exposure images are performed on the viewer
screen (for stitching combination position adjustment).
For details, refer to "7.8.3 Image combination".
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
Viewer screen (for stitching combination
position adjustment)
This is the screen for adjusting combined position of
images taken using Reina buckies with stitching expo-
sure.
(4)
(3)
(2)
(1)
(5)
(9)
(10)
(7)
(8)
(6)
Number Name Functions
(1)
Display layout
toggle button
Selects the screen display
format from 1×1, 2×1, and
2×2 display options.
(2) Overlap button
Switches and displays the
plate overlap sequence when
checking the combination
position. Switches to stan-
dard or reversed every time
it is pressed. The switching
is for display only and does
not aect the overlapping se-
quence during combination.
(3) Adjust button
Displays and hides the stitch-
ing adjustment panel.
(4) Proportion button
Changes the displayed trans-
mittance of the overlapped
portion. Pressing this button
sequentially changes the
transmittance. A modied
transmittance is not reected
in the output image.