7.5 Manual stitching exposure
2
Press [OFF] or the button of the output
size to be changed on the Split Output
Size dialog box.
The output size is changed.
HINT
•••••••••••••••••••••••••••••••••••••
When output of split images is enabled, a frame ap-
pears on the image to indicate the area of the split im-
age that will be output.
Split output frame
Trimming frame
The size of the trimming frame changes in accordance
with changes to the size of the image frame so that
images are automatically split into the selected section
size. Images can be split into a maximum of 3 sec-
tions.
If the split image is smaller than the trimming region,
the entire combined image may not completely fit
within the selected region.
If the conguration of the split images results in one
image, the split processing is not applied, and the split
output frame does not appear.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
7.5.2 Exposure
The manual stitching exposure method is as follows.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
For the stitching exposure images, make sure that the
combined parts are combined in the correct positions.
If there is gap between combined images, manually
adjust the image position of the combined parts using
the viewer screen (for stitching combination position
adjustment). It is also recommended that you arrange
a measure in the combined image part during expo-
sure to conrm the combined parts easier.
In stitching exposure, images can be combined when
all the following conditions are met.
All exposures are done using DR Detectors (CR
cassettes) of the same size
All exposures are done using an DR Detector or a
CR cassette
Perform exposure after standardizing the read resolution
When performing the manual stitching exposure in
AeroDR 3 1417 HD, only "Standard" (200µm) can be
selected for "Resolution".
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
After transitioning from the initial screen to
the exposure body part selection screen, se-
lect the Exam. Tag for stitching exposure.
Select 2 or 3 Exam. Tags for stitching exposure.
The selected Exam. Tags are displayed on the
Exam. Tags display area.