7.22 Other functions
Chapter 7
• When reading bar codes of the patient list, read
while keeping the distance from the bar codes to
approx. 5 to 10 cm. Sometimes it is not possible
to read outside this range.
*
Reading distance will vary depending on the
size of bar code and code system.
• Connect one PC to one Bluetooth bar code read-
er in tandem. Do not connect multiple units.
• During normal operation, if the PC is shut down,
connection of Bluetooth communication is discon-
nected, and the power of the bar code reader is
turned o 5 minutes later.
• The valid digit number cannot be set, depending
on the barcode reader used.
z
Magnetic card reader
• When a card is read while the screen saver is op-
erating, some data may be lost. Perform reading
after the screen saver is released.
• After reading magnetic data, always check that
the patient information displayed on the screen
matches the patient information written on the
magnetic card.
• Insert the magnetic card from the edge of the
magnetic card reader unit, and slide horizontally
to the opposite edge.
• While the magnetic data is being read, do not
perform input. Sometimes magnetic data gets
mixed with data input by input operation.
7.22.7 Addition of high preci-
sion monitor
An optional high precision monitor can be connected
to this device.
The following are precautionary items when a high
precision monitor is used.
• The image that has focus on the exposure screen
or the image displayed on the viewer screen is dis-
played. Unexposed images on the list screen, the
body part selection screen, and the exposure screen
are not displayed.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• Images cannot be manipulated on the high precision
monitor. Perform manipulation of images with this device.
• Since high precision monitors has a dierent aspect
ratio from that of the exposure screen and viewer
screen, when you enlarge, reduce, or move an image
on the exposure screen or viewer screen, an image of
a dierent area may be displayed on such a monitor.
Initial display
(Display magnication: 1)
Normal monitor
High precision monitor
Normal monitor High precision monitor
Enlarged image
(Display magnication: 1.3)
• The high-definition monitor does not display a high-
denition mammography
•••••••••••••••••••••••••••••••••••••••••••••••••••••