Chapter 6
6.6.1
Change of X-ray irradia-
tion information (results)
When it is necessary to change irradiation information
(results) after exposure, irradiation information (results)
can be changed with the following procedure.
1
Select an image to edit.
2
Press the [Exp.Info] tab.
3
Press the number part of the item you
want to edit.
6.6 Exposure Screen ([Exp.Info] Tab) Functions
This section describes the operation performed when
the [Exp.Info] tab is selected on the control panel of
the exposure screen. X-ray irradiation information can
be displayed on the image by the setting.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When linked to the X-ray device, editing can only be
made when X-ray irradiation results cannot be ac-
quired from the X-ray device.
• Risk of misidentication of patients may occur if modi-
cations to patient information are made. Be sure to
read "Chapter 1 Safety Precautions & Warnings".
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
•
When there are no X-ray irradiation results associated
with an exposed image, result values are not displayed.
•
Editing of X-ray irradiation results can be performed re-
gardless of the existence of links to the X-ray device.
•
Even if results were input before acquiring X-ray irradia-
tion results, they are overwritten with the irradiation re-
sults acquired from the X-ray device when the image is
read. If irradiation results could not be acquired from the
X-ray device, then the values are not overwritten.
•••••••••••••••••••••••••••••••••••••••••••••••••••••