392
10.2 Error codes and countermeasures
10.2.3 DR Detector errors
Error codes starting with "F" are indicating DR Detector errors.
Error code Displayed message Problems/countermeasures
F0100*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0101*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0200*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0201*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0202*
F0203*
[DR Detector name] version update error
If the panel after rebooting does not recover, contact a
service representative.
F0210*
F0211*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0300*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0400*
[DR Detector name] version update error
If the panel after rebooting does not recover, contact a
service representative.
F090**
F091**
F092**
F093**
F094**
[DR Detector name] radio error
The radio condition has transiently worsened.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
Wireless connection cannot be used, but continued
operation is possible with the wired connection.
F0H00*
F0H01*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0H02*
[DR Detector name] internal error.
If the panel after rebooting does not recover, contact a
service representative.
F0H03*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0H04*
F0H05*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0H06*
F0H07*
F0H08*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0H09*
F0H10*
F0H11*
F0H12*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0H16*
F0H17*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0I00*
F0I01*
F0I02*
F0I03*
[DR Detector name] power source error
If the panel after rebooting does not recover, contact a
service representative.
F0I04*
F0I05*
[DR Detector name] power source error
If the panel after rebooting does not recover, contact a
service representative.