393
10.2 Error codes and countermeasures
Chapter 10
Error code Displayed message Problems/countermeasures
F0I06*
F0I07*
[DR Detector name] capacitor error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0I08*
F0I09*
F0I10*
[DR Detector name] capacitor error
Can continue to use the panel. But, if the panel after
rebooting does not recover, contact a service repre-
sentative.
Lithium ion capacitor is degraded, or the power
consumption of the whole AeroDR SYSTEM has in-
creased.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
However, continued use is possible.
F0I11*
F0I12*
F0I13*
[DR Detector name] capacitor error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0I14*
[DR Detector name] power source error
If the panel after rebooting does not recover, contact a
service representative.
F0I15*
[DR Detector name] capacitor error
If the panel after rebooting does not recover, contact a
service representative.
F0I16*
[DR Detector name] power source error
If the panel after rebooting does not recover, contact a
service representative.
F0I17*
AeroDR Battery Charger error
If the panel after rebooting does not recover, contact a
service representative.
F0I18*
[DR Detector name] Power source error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector and AeroDR Battery Char-
ger/AeroDR Battery Charger2.
(2) Contact Konica Minolta technical representatives.
F0I19*
[DR Detector name] power source error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0I20*
[DR Detector name] power source error
Remove cable or cradle from the panel.
If the panel does not recover after a while, contact a
service representative.
(1) Remove the DR Detector from the AeroDR Battery
Charger/AeroDR Battery Charger2 or the wired
connection cable.
(2) Contact Konica Minolta technical representatives.
F0I21*
[DR Detector name] capacitor error
Contact a service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0I22*
[DR Detector name] power source error
If the panel after rebooting does not recover, contact a
service representative.
F0I23*
[DR Detector name] power source error
The AeroDR panel cannot be charging because of high
temperature inside the panel.
You can charge it again after cooling down.
(1) Move the DR Detector away from an electrical
carpet, a location exposed to direct sunlight, or a
place where heat can build up, then wait until the
temperature falls.
If an error occurs, charging will not be possible.
However, it will be possible to carry out exposures.
(2) If the problem persists, contact Konica Minolta tech-
nical representatives.
F0I24*
F0I25*
[DR Detector name] capacitor error
If the panel after rebooting does not recover, contact a
service representative.
(1) Restart the DR Detector.
(2) Contact Konica Minolta technical representatives.
F0I26*
[DR Detector name] internal error
If the panel after rebooting does not recover, contact a
service representative.
F0I27*
[DR Detector name] capacitor error
If the panel after rebooting does not recover, contact a
service representative.
F0I28*
AeroDR Battery Charger error
If the cradle after rebooting does not recover, contact a
service representative.
(1) Restart the AeroDR Battery Charger/AeroDR Bat-
tery Charger2.
(2) Contact Konica Minolta technical representatives.
Continued on the next page