9.1 Various problems and countermeasures
377
Chapter 9
6
The confirmation screen is displayed on
this unit connected to the same system.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
After the switchover of JM, contact Konica Minolta
technical representatives as soon as possible.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
Switching to backup JM may take about 1 minute.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
9.1.2 Reacquisition of images
from the DR Detector
Reacquire the images when images cannot be re-
ceived from the DR Detector after the exposure due to
reasons such as communication errors, etc.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Reacquire images that have not been transferred us-
ing this device, in which a communication error has
occurred.
If you end your examination while an image that has
not been transferred is left in the DR Detector, you will
not be able to acquire the image again.
When images that have not been transferred remain in
the DR Detector, exposure cannot be performed with
that DR Detector.
If an image cannot be received using the Auto stitch-
ing exposure with DR Detector, cancel the image
reacquisition and do not perform any changes to the
exposure device.
In the event of a communication error and depending
on the timing at which the error has occurred, images
may not be displayed and the DR Detector may not
contain any images that have not been transferred
(the same status as that observed when no exposures
have been carried out). In such a case, a message
may be displayed. Suspend the study and display the
exposure screen from the list screen. This operation
may successfully recover the images.
Non-transferred images need to be reacquired be-
cause the One Shot stitching exposure conducts the
combining and displaying images after acquiring all
the images.
Delete the non-transferred images from all the DR
panels used in the exposure, to delete the non-
transferred images by the reacquiring of the One Shot
stitching exposure.
(a) Press [OK] when a message "Unreceived image
of One Shot Stitching exposure is deleted. Unre-
ceived image of same exposure is present in other
panels." appeared.
(b) Press the framed device icon.
(c) Press [OK] when a conrmation screen appeared.
(d) Press [OK] when the conrmation screen appeared
again.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Other manuals for Konica Minolta CS-7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CS-7 and is the answer not in the manual?

Konica Minolta CS-7 Specifications

General IconGeneral
BrandKonica Minolta
ModelCS-7
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals