EasyManua.ls Logo

Konica Minolta CS-7 - Exposure

Konica Minolta CS-7
424 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
212
7.2 Exposure with AeroDR Portable RF Unit2
7.2.5 Exposure
Prepare for exposure with the AeroDR Portable RF
Unit2 according to the following procedure.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Always check the patient information and exposure
conditions before exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Conrm that each device is ready for ex-
posure.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Check the wireless connection.
If a wired connection is established between this de-
vice and the AeroDR Portable RF Unit2, disconnect
the Ethernet cable that connects this device and the
hub.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Perform exposure.
Congure the mobile X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
The exposed image is displayed in the image
display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When performing an exposure, stay away from the pa-
tient as far as possible before pressing the exposure
switch.
Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
If an image is displayed when an exposure is not per-
formed, this may be a detection error due to noises,
etc. Perform rejected image (deletion of images) and
repeat exposure.
If you have performed an exposure when "Ready" was
not displayed, repeat an exposure after approximately
one minute.
When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD. Selecting the AeroDR 3 1417HD
icon allows the exposure time to be selected.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
The selected Exam. Tag is added to the Exam.
Tags display area. By repeating from step 1,
Exam. Tags can be added.
4
When nished selecting/adding the Exam.
Tags, press [Perform].
The exposure screen is displayed.
HINT
•••••••••••••••••••••••••••••••••••••
When only reserving an examination, after adding the
Exam. Tag, press [Schedule]. After reservation is per-
formed, the initial screen is displayed.
Returning to the initial screen by pressing [Schedule]
does not set study date and time of image supplemen-
tal information automatically.
It is also possible to register multiple exposure methods
in 1 Exam. Tag. Register these methods when install-
ing.
If you have set to register the Exam. Tags automati-
cally, the body part selection screen is not displayed.
The display transitions to the exposure screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents