7.3 Exposure with AeroDR Portable UF Unit
Chapter 7
7.3.2 Exposure preparations
Make the following preparations when exposing using
the AeroDR Portable UF Unit. The following procedure
is for an example in which a wired connection is in op-
eration between this device and a higher system.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When this device is being used in the AeroDR Por-
table UF Unit, charge it when a notification that the
battery level is low is displayed.
• Check the settings for the AeroSync button, and initial
detection level on the system monitor screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Register the DR Detector in this device
that is to be used in AeroDR Portable UF
Unit.
• For operation of the DR Detector, refer to
"AeroDR Operation Manual".
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When an DR Detector which does not support Aero
Sync mode is used, an error message appears and
registration will not be permitted.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• Conrm that the outputs of host, RIS-OUT and printer
are [ON] on the system monitor screen ([System Moni-
tor] tab).
In case of the host
The device switches between [ON] and [OFF] every
time [Output] is pressed.
• When using a RIS-OUT or printer, conrm the output is
set to [ON] using the same procedure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Import the examination information to be
used for the exposure using the portable
X-ray device.
• Conrm that an Ethernet cable is connected to
this device.
• Press [Refresh] on the list screen on this device.
• The latest examination information is imported
from the higher system by pressing the [Refresh]
button.
HINT
•••••••••••••••••••••••••••••••••••••
• To input examination information manually, change the
screen to the patient registration screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Remove the Ethernet cable connected to
the DR Detector registration side hub from
this device.
HINT
•••••••••••••••••••••••••••••••••••••
• When this device is connected to the AeroDR Portable
UF Unit through wired connection, remove the Ether-
net cable connected to the DR Detector registration
side hub from the portable X-ray device side hub.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
4
Unplug the power cable from the outlet
when charging this device.
5
Complete the exposure preparations.
• Confirm that the AeroDR Portable UF Unit is
shut down to complete the exposure and move-
ment preparations.
HINT
•••••••••••••••••••••••••••••••••••••
• It is all right to keep power of this device and the DR
Detector on.
•••••••••••••••••••••••••••••••••••••••••••••••••••••