236
7.4 Exposure with Unitless
7.4.3 Exposure
Prepare for exposure using unitless according to the
following procedure. The procedure varies depending
on the setting of the [AeroSync Button] on the system
monitor screen.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• Always check the patient information and exposure
conditions before exposure.
• Do not disconnect the Ethernet cable while the data is
being transmitted.
• The status of communication with the host is displayed
on the [System] button. For detail, refer to "7.4.1
Screen structure".
• Available X-ray detection level varies depending on
the DR Detector model.
• Use the PDA at a distance within 10 m of this device
and the DR Detector.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• The procedure is explained with the use of the Aero-
DR 3 1417HD as an example.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
When the [AeroSync Button] is OFF
1
After arriving to the imaging location, con-
rm that the icon for the DR Detector used
for imaging appears in the application bar.
2
Select the examination.
3
Press [Perform].
• The exposure screen is displayed.
• A message "Ready" is displayed in the mes-
sage/button area, then the screen is locked.
• To change the X-ray detection level, press
[Pause] to enable modification. After changing
the detection level, press [Restart], then the
message "Ready" is displayed.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• A malfunction may be caused by a shock to the DR
Detector. In such cases, set [AeroSync Button] to ON
on the system monitor screen.
• To prevent incorrect operations, always check that
"Ready" is displayed.
• Conditions for changing the detection level vary de-
pending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• When importing examination information from the
higher system and when exposure body part information
is not imported, select the exposure body part. When
pressing [Perform] on the body part selection screen, the
exposure screen is displayed.
• When you move the focus in the image display area on
the exposure resumption waiting screen to capture an im-
age of another body part, not the ready-to-expose screen
but the exposure resumption waiting screen is displayed.
• Refer to "7.4 Exposure with Unitless" for information on
the X-ray detection levels that can be set for exposures.
•••••••••••••••••••••••••••••••••••••••••••••••••••••