236
7.4 Exposure with Unitless
7.4.3 Exposure
Prepare for exposure using unitless according to the
following procedure. The procedure varies depending
on the setting of the [AeroSync Button] on the system
monitor screen.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Always check the patient information and exposure
conditions before exposure.
Do not disconnect the Ethernet cable while the data is
being transmitted.
The status of communication with the host is displayed
on the [System] button. For detail, refer to "7.4.1
Screen structure".
Available X-ray detection level varies depending on
the DR Detector model.
Use the PDA at a distance within 10 m of this device
and the DR Detector.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
The procedure is explained with the use of the Aero-
DR 3 1417HD as an example.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
When the [AeroSync Button] is OFF
1
After arriving to the imaging location, con-
rm that the icon for the DR Detector used
for imaging appears in the application bar.
2
Select the examination.
3
Press [Perform].
The exposure screen is displayed.
A message "Ready" is displayed in the mes-
sage/button area, then the screen is locked.
To change the X-ray detection level, press
[Pause] to enable modification. After changing
the detection level, press [Restart], then the
message "Ready" is displayed.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
A malfunction may be caused by a shock to the DR
Detector. In such cases, set [AeroSync Button] to ON
on the system monitor screen.
To prevent incorrect operations, always check that
"Ready" is displayed.
Conditions for changing the detection level vary de-
pending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
When importing examination information from the
higher system and when exposure body part information
is not imported, select the exposure body part. When
pressing [Perform] on the body part selection screen, the
exposure screen is displayed.
When you move the focus in the image display area on
the exposure resumption waiting screen to capture an im-
age of another body part, not the ready-to-expose screen
but the exposure resumption waiting screen is displayed.
Refer to "7.4 Exposure with Unitless" for information on
the X-ray detection levels that can be set for exposures.
•••••••••••••••••••••••••••••••••••••••••••••••••••••