7.4 Exposure with Unitless
Chapter 7
4
Perform exposure.
Congure the portable X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
As soon as the exposed image is transferred,
the image is displayed in the image display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When performing an exposure, stay away from the pa-
tient as far as possible before pressing the exposure
switch.
When the "Ready" is displayed, immediately perform
exposure.
Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
If image acquisition started without performing ex-
posure, this may result in incorrect detection due to
noise, etc. Perform reject (image deletion) and repeat
exposure.
If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con-
ditions and repeat exposure.
When exposure is done while "Ready" is not dis-
played, have an interval of 1 minute or more before
repeating exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
5
Adjust the image.
After exposure, adjust the image as required.
6
Press [ ].
The exposure is completed, and the display re-
turns to the initial screen.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When consolidating image adjustment, press [
] to
pause the examination and return to the initial screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
Image adjustment can be done on the portable de-
vice's [Image Proc] tab. However, if a more advanced
image adjustment is required, perform image adjust-
ment on the [Image Proc] tab or the viewer screen.
For details on the viewer screen, refer to "6.7 Func-
tions of the viewer screen".
If you have to move to the imaging location, perform
the procedure again from step 1 after arriving to the
imaging location.
It is all right to keep power of this device and the DR
Detector on.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
When the [AeroSync Button] is ON
1
After arriving to the imaging location, conrm
that the icon for the DR Detector used for im-
aging appears in the application bar.
2
Select the examination.
3
Press [Perform].