7.4 Exposure with Unitless
• The exposure screen is displayed.
• A message "Waiting for permission" is displayed
in the image display area.
HINT
•••••••••••••••••••••••••••••••••••••
• If the information about the body part is not imported
while importing examination information from the high-
er system, select the body part to be imaged. Then
press [Perform] on the body part selection screen to
display the exposure screen.
• When the AeroDR power management level is set to
[3], [4] or [5], "Waiting to start shooting" is displayed
instead of "Waiting for permission".
•••••••••••••••••••••••••••••••••••••••••••••••••••••
4
After positioning the patient, check the X-
ray detection level, then press [AeroSync].
• If the control panel does not switch to
[AeroSync], press the [AeroSync] tab.
• Change the X-ray detection level as needed.
• After checking the X-ray detection level, press
[AeroSync].
• A message "Ready" is displayed in the mes-
sage/button area. The screen is locked.
• The screen is locked. To return to the previous
screen to change the position of the patient,
etc., press [Cancel].
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• To prevent incorrect operations, always press
[AeroSync] and check that "Ready" is displayed.
• Conditions for changing the detection level vary
depending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
• When the AeroDR power management level is set to
[3], [4] or [5], it may take some time until the exposure
becomes ready after you have pressed [Aero Sync]. In
this case, [Standby] button instead of [Cancel] button
is displayed in the message/button area.
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• Refer to "7.4 Exposure with Unitless" for information
on the X-ray detection levels that can be set for expo-
sures.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
5
Perform exposure.
• Congure the portable X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
•
As soon as the exposed image is transferred,
the image is displayed in the image display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
When performing an exposure, stay away from the patient
as far as possible before pressing the exposure switch.
• When the "Ready" is displayed, immediately perform
exposure.
• Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
• If image acquisition started without performing exposure,
this may result in incorrect detection due to noise, etc.
Perform reject (image deletion) and repeat exposure.
• If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con-
ditions and repeat exposure.
• When exposure is done while "Ready" is not dis-
played, have an interval of 1 minute or more before
repeating exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••