271
7.8 CR stitching cassette exposure
Chapter 7
1
Focus on Exam. Tags to register the CR
stitching cassette.
2
Read the barcode of the CR stitching cas-
sette with a barcode reader.
• The barcode number read is displayed in the
Exam. Tag display area. This completes the
read preparation for REGIUS.
3
Perform exposure with the registered CR
stitching cassette.
•
Configure the X-ray device settings and check
the exposure setting before making an exposure.
Press the exposure switch
4
Set the REGIUS plate on the read cassette
and read by REGIUS.
• The image read from the barcode on the CR
stitching cassette will be displayed.
7.8.3 Image combination
Images exposed by CR stitching cassette can be com-
bined automatically or manually. The procedure below
explains an example of combining 3 images.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• Adjust the combination position by using the markers
as supplementary standards in addition to body parts
of the subject.
It is also recommended that you arrange the measure
in the combined image part during exposure to sim-
plify the conrmation of the combined parts.
• In stitching exposure, when plates overlap and are ar-
ranged so that there is no image loss in each plate's
boundaries, there might be a slight gap in the marker
position because of a minute dierence in magnica-
tion. After taking this into account, conrm the combi-
nation position through the marker.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Select the images to be combined.
• Select the images to be combined from the or-
der list and press [ ] or [ ].
• The selected images are displayed in the com-
bined image display area.
• If you pressed [ ], continuously select the
associated images (second and third).
HINT
•••••••••••••••••••••••••••••••••••••
• [ ] and [ ] can be pressed only when the
Exam. Tag for stitching exposure has been selected.
•••••••••••••••••••••••••••••••••••••••••••••••••••••