EasyManua.ls Logo

Konica Minolta CS-7 - Page 203

Konica Minolta CS-7
424 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
203
7.1 Exposure with AeroDR Portable RF Unit
Chapter 7
4
Perform exposure.
Congure the portable X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
The exposed image is displayed in the image
display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When performing an exposure, stay away from the pa-
tient as far as possible before pressing the exposure
switch.
Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
If an image is displayed when an exposure is not per-
formed, this may be a detection error due to noises,
etc. Perform rejected image (deletion of images) and
repeat exposure.
If you have performed an exposure when "Ready" was
not displayed, repeat an exposure after approximately
one minute.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
5
Adjust the image.
After exposure, adjust the image as required.
6
Press [ ].
The exposure is completed, and the display re-
turns to the initial screen.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When consolidating image adjustment, press [
] to
pause the examination and return to the initial screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
7
When exposure is nished, turn the Aero-
DR Portable RF Unit o󰀨.
Turn off the power switch of the AeroDR Por-
table RF Unit, and check that the power LED
(green/orange) is o󰀨.
For shutdown operation of the AeroDR Portable
RF Unit, refer to the "AeroDR Portable RF Unit
Operation Manual".
HINT
•••••••••••••••••••••••••••••••••••••
Image adjustment can be done on the exposure
screen. However, if a more advanced image adjust-
ment is required, perform image adjustment on the
viewer screen. For details on the viewer screen, refer
to "6.7 Functions of the viewer screen".
When moving the exposure location, turn off the
AeroDR Portable RF Unit. After moving the exposure
location, perform the procedure again from step 1.
It is all right to keep power of this device and the DR
Detector on.
•••••••••••••••••••••••••••••••••••••••••••••••••••••