232
7.4 Exposure with Unitless
z
Exposure screen
This is the control panel used for exposing with unitless.
(2)
(1)
Number
Name Functions
(1) X-ray detection level switching area Select the X-ray detection level during exposure.
(2) Message/Button area
Messages and buttons are displayed in this area depending on the opera-
tion status.
When the [AeroSync Button] is set to ON on the system monitor
screen
Exposure permission waiting screen
The device is waiting for permission for exposure. The
screen changes to the ready-to-expose screen if you
press [AeroSync].
Ready-to-expose screen
The screen is locked as it is ready to be captured. To
cancel exposure, press [Cancel], then the screen re-
turns to the exposure permission waiting screen.
When the [AeroSync Button] is set to OFF on the system monitor
screen
Ready-to-expose screen
The screen is locked as it is ready to be captured. To
suspend exposure, press [Pause], then the screen returns
to the exposure permission waiting screen.
Exposure resumption waiting screen
The device is waiting for exposure to be resumed. The
screen changes to the ready-to-expose screen if you
press [Restart].
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
If the icon indicating that unitless imaging is not available appears, restart the operating system (OS) for this device. Imaging
cannot be performed until the OS is restarted.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••