7.10 Mammogram exposure
Chapter 7
7
Expose using the registered Mammogra-
phy cassette.
• Make sure the image display areas to be ex-
posed are focused.
• Set up the X-ray device before exposure. Per-
form exposure after checking the exposure set-
tings.
Press the exposure switch
8
Read the Mammography cassette with RE-
GIUS.
• The image read from the barcode on the Mam-
mography cassette will be displayed.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• In the case of X-ray link, make sure exposure settings
are displayed for the image after exposure.
• Confirm that additional information such as markers
and stamps is not superimposed on the breast area.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• Overlay settings (LR, CC, MLO, etc.) are made on the
[Overlay] tab (host or printer). For information on the
overlay setting method, refer to "6.4.22 Overlay".
• If the image needs revision (chest wall blackening
process, etc.) when checking the image by pressing
[Preview] of overlay, adjust the image on the viewer
screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
9
Perform the second exposure or more.
• Repeat from step 4 to perform the second expo-
sure or more.
10
Press [ ] for the rst image.
11
Press [ ] for the second image.
• The breast position adjustment screen is dis-
played. Two images are displayed on the right
and left, and the vertical position of them is au-
tomatically adjusted.
HINT
•••••••••••••••••••••••••••••••••••••
• When the two images are transposed with each other,
press [Swap] to replace them with each other.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
12
Fine-tune the position of the right and left
images if necessary.