29
Chapter 1
1.2 Safety precautions
CAUTION
Precautions when sending exposure setting
• When sending exposure setting to an external X-
ray device, check the exposure conditions on the X-
ray device. Be especially careful of X-ray exposure
settings when performing an exposure on a small
child.
Precautions when delivering an image from
this device to other devices
• Conrm that additional information such as set an-
notation is correct.
Precautions for exposure with the DR Detector
• Before exposure, conrm on the screen of this de-
vice that the DR Detector is registered in the expo-
sure room. Also, perform exposure after conrming
that the DR Detector to be used for exposure has
been selected.
• Do not leave the DR Detector in the vicinity of a
heating device such as a heated carpet etc.
Precaution on exposure dose
• After exposure, S values are displayed in the im-
age display areas of this device. Check that the S
values indicate the appropriate dosage for exposed
images.
Precautions when performing exposure in Aero
Sync mode
• Confirm that the device is ready for exposure
through its display before performing exposure.
• Conrm that a conrmation is sounded from the de-
vice after the start of exposure.
• Do not let the DR Detector vibrate or receive shock
until the preview image is displayed.
• Conrm that the DR Detector’s battery level is full
before performing exposure.
• Use within the range of the pre-checked exposure
conditions.
• Be aware that the exposure procedure must be
changed in some cases when the settings are
changed for the Aero Sync button or initial detection
level. Be sure to press the Aero Sync button after
changing the Aero Sync button setting from OFF to
ON.
• If the Aero Sync button or initial detection level has
not been set appropriately, a malfunction may occur
due to the DR Detector being subjected to impact.
Make sure values are changed appropriately.