Protocol analysis
R&S
®
RTP
608User Manual 1337.9952.02 ─ 12
"Word Count Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:MILStd:CMD:CCONdition on page 1690
TRIGger<m>:MILStd:CMD:CMAX on page 1691
TRIGger<m>:MILStd:CMD:CMIN on page 1691
T/R (Transmit/receive)
Toggles the data direction of the selected command: 1 (transmit), 0 (receive), or X
(either).
Remote command:
TRIGger<m>:MILStd:CMD:TR on page 1692
Status Flags
Specifies the values of the status flags. You can use the following characters: 1; 0; or X
(do not care).
For details, see "Status Word" on page 598.
Remote command:
TRIGger<m>:MILStd:STATus:BCReceived on page 1696
TRIGger<m>:MILStd:STATus:BUSY on page 1696
TRIGger<m>:MILStd:STATus:DBCaccept on page 1696
TRIGger<m>:MILStd:STATus:INSTrument on page 1697
TRIGger<m>:MILStd:STATus:MERRor on page 1697
TRIGger<m>:MILStd:STATus:SREQuest on page 1697
TRIGger<m>:MILStd:STATus:SUBSystem on page 1697
TRIGger<m>:MILStd:STATus:TERMinal on page 1697
Error Condition
Specify the error conditions to be triggered on.
Synchronization Error ← Error Condition
Triggers if a sync impulse does not fulfill the technical requirements or when the trans-
mission is not valid.
Remote command:
TRIGger<m>:MILStd:ERRor:SYNC on page 1695
Manchester Coding Error ← Error Condition
Triggers if there is an error in the Manchester coding of the signal.
Remote command:
TRIGger<m>:MILStd:ERRor:MANChester on page 1695
Parity Error ← Error Condition
Checks the parity of every word and triggers if the parity is even.
Remote command:
TRIGger<m>:MILStd:ERRor:PARity on page 1695