Protocol analysis
R&S
®
RTP
777User Manual 1337.9952.02 ─ 12
Threshold High
Sets the threshold high of the signal, if "Signal type" > "Differential".
Remote command: 
BUS<m>:EBTB:DIFFerential:THRHigh on page 1975
Threshold Low
Sets the threshold low of the signal, if "Signal type" > "Differential".
Remote command: 
BUS<m>:EBTB:DIFFerential:THRLow on page 1975
Threshold D+
Sets the high threshold (D+) of the signal, if "Signal type" > "Single ended".
Remote command: 
BUS<m>:EBTB:DPLus:THReshold on page 1977
Threshold D-
Sets the high threshold (D-) of the signal, if "Signal type" > "Single ended".
Remote command: 
BUS<m>:EBTB:DMINus:THReshold on page 1976
Find thresholds
Executes the measurement of reference levels and sets the thresholds to the middle
reference level of the measured amplitude.
Remote command: 
BUS<m>:SETReflevels on page 1548
Bitrate
Sets the number of transmitted bits per second.
Remote command: 
BUS<m>:EBTB:BITRate on page 1974
Estimate Bitrate
Starts a software algorithm for the automatic determination of the bitrate. The value is
displayed in the "Estimated Bitrate".
Remote command: 
BUS<m>:EBTB:BITDetermi on page 1978
BUS<m>:EBTB:EBTRate? on page 1975
Show thresholds
If enabled, the threshold levels are displayed in the diagram.
Remote command: 
BUS<m>:THReshold on page 1549
13.16.2.2  Advanced settings
Access: [Protocol] > "Advanced" tab.
8b/10b (option R&S
 RTP-K52)