EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 238

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Internal Baseband Source
R&S
®
SMBV100B
236User Manual 1178.4460.02 ─ 03
Delay Start.................................................................................................................. 237
Delay Step...................................................................................................................237
Input Waveform File.................................................................................................... 237
Apply Assistant Settings..............................................................................................237
Carrier State
Switches the carriers in the range "Carrier Start" to "Carrier Stop" on/off.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STATe on page 769
Carrier Start/Start
Defines the start/stop index of the carrier range to which the assistant settings are
intended to apply.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt on page 770
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP on page 770
Gain Start
Sets the gain of the carrier marked by "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt]
on page 770
Gain Step
Sets the step width that is used to increment the gain.
The resulting carrier gain in the carrier table equals:
Gain =
"Gain Start" + n*"Gain Step".
Where n is a value between 0 and ("Carrier Stop" - "Carrier Start").
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
on page 770
Phase Start
Sets the phase of the carrier marked by "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe[:STARt]
on page 771
Phase Step
Sets the step width that is used to increment the phase.
The resulting phase in the carrier table equals:
Phase =
"Phase Start" + n*"Phase Step"
Where n is a value between 0 and ("Carrier Stop" - "Carrier Start").
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
on page 770
Generating Multi-Carrier Signals

Table of Contents

Related product manuals