Remote Control Commands
R&S
®
SMBV100B
769User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:DELay <Delay>
Sets the start delay of the selected carrier.
Parameters:
<Delay> float
Range: 0 to 1
Increment: 1E-9
*RST: 0
Default unit: s
Example:
BB:ARB:MCAR:CARR15:DEL 5us
sets a start delay of 50 us for carrier 15.
Manual operation: See "Delay" on page 234
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:FILE <File>
Selects the file with I/Q data to be modulated onto the selected carrier.
Parameters:
<File> <file name>
Example: See Example "Generating a multi carrier signal" on page 744
Manual operation: See "File" on page 234
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:CONFlict?
Queries carrier conflicts. A conflict arises when the carriers overlap.
Return values:
<Conflict> 0 | 1 | OFF | ON
0
No conflict
*RST: 0
Example:
BB:ARB:MCAR:CARR:CONF?
Usage: Query only
Manual operation: See "!!!" on page 235
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STATe <State>
Switches all the carriers in the selected carrier range on or off.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
BB:ARB:MCAR:EDIT:CARR:STAT ON
Sets all the carriers in the carrier range to ON.
SOURce Subsystem